CELL-DYN
®
3200 System Operator’s Manual
2-5
9140181K—July 2002
Section 2
Installation Procedures and Special Requirements
Tubing Installation
Reagent and Waste Tubing
1. Locate the reagent tubing in the Accessory Kit.
2. Inspect each length of tubing carefully for damage or cracks.
3. Attach the non-weighted end of the tubing with the
Purple WBC Lyse
label
to the
Purple
Connector on the right side of the Rear Panel. (Refer to Figure
2.1.) This reagent is for the WBC dilution. Wipe the outside of the tubing
with a damp lint-free pad (such as DYN-A-WIPE) and place the weighted
end into the container of CELL-DYN WBC Lyse. Secure the cap.
Place the
container on the same level as or lower than the unit.
4. Attach the non-weighted end of the tubing with the
Red Diluent/Sheath
label to the
Red
Connector. This reagent is for the RBC dilution and for
rinsing the flow system. Wipe the outside of the tubing with a damp lint-free
pad (such as DYN-A-WIPE) and place the weighted end into the container of
CELL-DYN Diluent/Sheath. Secure the cap.
Place the container on the same
level as or lower than the unit.
5. Attach the non-weighted end of the tubing with the
Blue HGB Lyse
label to
the
Blue
Connector. This reagent is for the HGB dilution. Wipe the outside
of the tubing with a damp lint-free pad (such as DYN-A-WIPE) and place the
weighted end into the container of CELL-DYN HGB Lyse.
Secure the cap.
Place the container on the same level as or lower than the unit
.
6. Attach the
Waste Outlet Tubing
to the
:DVWH
Outlet Connector. Place the
end of the tubing with the cap and sensor into the waste collection container.
Ensure that the waste collection container is adequately labeled. Secure the
cap, or remove the cap from the tubing and place the tubing into a drain
suitable for collection of waste with
possible biological and chemical hazard
.
Be sure that the tubing is secured to the drain hole.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...