CELL-DYN
®
3200 System Operator’s Manual
11-17
9140181K—July 2002
Section 11
Quality Control
Running Controls
Abbott recommends using CELL-DYN Control materials for performing Quality
Control checks on the CELL-DYN 3200 System. Run a minimum of two levels of
control at the beginning of each eight hours of operation prior to running patient
samples. In addition, run controls after the following events:
• After calibration (confirmatory step).
• After Daily Start-Up procedures are completed.
• After a reagent lot number change.
• After maintenance, a service call, or component replacement.
• After detection of an unusual trend or shift in patient results.
• In accordance with the laboratory's Quality Control protocol.
• According to regulatory requirements.
Controls should be run in both operating modes, (Open or Closed). Always do the
following:
• If the System has been idle for 15 minutes or more, run a background
immediately prior to running a control specimen.
• Follow the proper warming and mixing procedures previously described,
including those located on the control package insert.
• Run control samples for each measured parameter in the same manner as
patient samples.
• Verify that control results are within the laboratory's acceptable limits.
• If the control results fall within acceptable limits, review the data for shifts or
trends, record the results, and begin to process patient samples.
• If one or more result falls outside the laboratory's acceptable limits, review
Troubleshooting and Diagnostics
;
. Perform suggested maintenance steps for the
parameter that is out-of-range, then try using another vial from the same lot.
If the problem persists, contact the Abbott Hematology Customer Support
Center.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...