4-10
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Performance Characteristics and Specifications
Performance Specifications
Section 4
Reticulocyte % measurement carryover, as shown in Table 4.7, is calculated using
actual Listmode counts, rather than using the Reticulocyte percent. It is calculated
using the following formula:
Reticulocyte carryover is determined on fresh whole blood samples with RBC in
the range 4.0–6.0 M/µL. Retic Background Count is reported on the Retic Run
Results Screen, while Retic Listmode can be found in DIAGNOSTICS
→
RETIC
RAW DATA.
Precision
Precision is the measurement of how closely the instrument is able to reproduce the
same results from repeated runs of the same sample.
Table 4.8 and Table 4.9 present the limits of acceptable precision for specimens run
in both the Open and Closed modes. The stated CV% in these tables represents the
instrument’s precision at a 95% confidence level from N=20 replicate runs based
on the same sample.
NOTE:
95% confidence limit means that at least 19 of the 20 determinations meet
this limit.
NOTE:
N=20 is based on one whole blood sample aliquotted into 4 tubes, with
each tube run 5 times in Closed mode.
Fresh whole blood samples that are used to verify precision specifications should
have results that fall within the laboratory’s normal range. These samples should
not display any of the following suspect parameter flags:
WBC
LRI
URI
LURI
The precision values listed in this section are applicable to Patient, Fragile WBC,
Resistant RBC (Open mode only), and Reticulocyte RBC (Open Mode Only)
specimen types.
Table 4.7
Carryover for Reticulocyte
Parameter
Specimen Range
Carryover
RETIC %
1.8%–20%
with 30,000 Listmode events
< 0.5%
% Carryover =
Background
1
-
Background
3
Retic Listmode
3
- Background Count
3
X
100
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...