CELL-DYN
®
3200 System Operator’s Manual
6-11
9140181H—October 2001
Section 6
Calibration Procedures
8.______
Verify the Closed Mode precision as follows:
a. For the CS model: Use the same sample as in the Open Mode. Select an
empty QC file in the
SPECIMEN TYPE
menu. Make sure Closed Mode is
selected. Run the sample 10 times.
b. For the SL model: Use the same sample as in the Open Mode. Aliquot the
blood in equal volumes into 10 5-mL tubes (each tube should contain a
minimum of 1 mL of sample) containing no anticoagulant. Select an
empty QC file in the
SPECIMEN TYPE
menu. Make sure Closed Mode is
selected. Place the tubes in a rack, place the rack in the “loading”
position, and press the
[START LOADER]
key.
c. When all the samples have been processed on either the CS or SL model,
record the CVs below or attach a file printout to this document.
PARAMETER
CV% LIMIT
CV%
WOC
<
2.7%
________
NOC
<
2.7%
________
RBC
<
1.5%
________
HGB
<
1.0%
________
MCV
<
1.0%
________
PLT
<
4.0%
________
9.______
If any problems are detected during the procedures outlined above, document them on the form on
the following page. Make copies of this form as necessary.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...