10-68
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
Loader Fault 48: tube stuck in position 4
/BL (Continued)
2. The tube position sensor is dirty.
2. Remove the Loader Cover. Using a
DYN-A-WIPE pad or lint-free
wipe moistened with lens cleaner or
deionized water, clean the tube
sensors (the tube position 4 sensor
is on the left). Dry the sensors
thoroughly. Then, reinstall the
Loader Cover, reset the racks, and
reset the Sample Loader.
NOTE:
For the location of the tube
sensors, refer to the figure
in
3. A failure of the sensor or related
electronics has occurred.
3. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Tower Fault 49: tower cover open
/BL
The circuit formed when the Tower Cover is in place has been broken.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The Tower Cover has been removed
or is not seated properly.
1. Reinstall or reset the Tower cover.
Make sure the cover is held in
position by the magnets on the
instrument frame. Then reset the
racks and reset the Sample Loader
by pressing the following keys in
order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
2. A failure of the sensor or related
electronics has occurred.
2. Reset the racks and reset the
Sample Loader. If the message
appears repeatedly, contact your
Abbott Hematology Customer
Support Center.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...