CELL-DYN
®
3200 System Operator’s Manual
6-19
9140181H—October 2001
Section 6
Calibration Procedures
Auto-Cal Method
Overview
The Auto-Cal program provides an automated calibration method that prepares the
CELL-DYN 3200 System for calibration, calculates new calibration factors, and
calibrates the instrument.
NOTE:
Always complete the Pre-Calibration procedures before beginning any
calibration.
Methodology
The instrument should be calibrated in both Open and Closed Modes using either
the CELL-DYN Calibrator or normal, fresh whole blood.
NOTE:
Auto-Calibration is not available in the Closed Mode on the SL model. To
calibrate Closed Mode, see
Calibration Procedure — Mode to Mode
in
subsection
Enter Factor Method
.
The same reference values and the same samples should be used when calibrating
both modes. Either mode may be calibrated first, but it is recommended to calibrate
the primary mode first.
NOTE:
Calibrator is the preferred material for calibrating the CELL-DYN 3200
System.
When samples are run, Auto-Cal does the following:
1. Accepts up to ten consecutive sample runs for either calibrator or fresh whole
blood samples.
2. Compares sample results against internal precision and reference checks,
highlighting results that fail these checks.
3. Calculates the new calibration factors (Mean Factors) and Factor % Diff
values.
4. Compares the Factor % Diff values to ranges in an internal table to determine
which parameters require calibration.
NOTE:
These ranges are shown in the Calibration Range Criteria
worksheets (Worksheet 3 for Open Mode and Worksheet 6 for
Closed Mode) in
Manual Calibration Worksheets
at the end of this
section.
5. Highlights Factor % Diff values which require calibration or which are over-
limit.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...