CELL-DYN
®
3200 System Operator’s Manual
5-91
9140181H—October 2001
Section 5
Operating Instructions
Closed Mode Analysis
The Closed Sampler mode on CELL-DYN 3200SL models aspirates the sample
from a closed collection tube that has been placed in a Sample Loader rack and
loaded into the Sample Loader.
CLOSED SAMPLER
is displayed in the upper right
corner of the
RUN
screen when this mode is selected. The wash block moves down
to the end of the open sample aspiration probe when this mode is selected.
Closed Mode Procedure
1. If necessary, in the
RUN
screen press
[CHANGE SAMPLER]
to select the Closed
Mode.
2. Be sure that the green READY light on the Analyzer Status Indicator Panel
is illuminated and
READY
is displayed in the Status Box on the
RUN
screen.
3. Place the well mixed samples in the Sample Loader racks.
NOTE:
The racks and tube positions are identified by the bar code label on
the rack and indicated as RxTx. These numbers appear as the
Sample ID number if bar code labels are not used. (If the Work List
is used, the Sample ID number is taken from it. Refer to
Using the
Work List
later in this section for more information.)
4. Place the racks in the Sample Loader to the right of the Tower Cover with the
slotted side (with bar code labels) facing the operator.
5. Verify the Tower Cover is in place.
NOTE:
The Sample Loader will not operate without the Tower Cover in
place.
6. Press the
[START LOADER]
key on the
RUN
screen.
7. The Sample Loader automatically processes all the samples. Processing stops
when the last rack is finished or the
[STOP LOADER]
key is pressed. The last
rack is finished when it has moved completely to the unload side of the
Sample Loader.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...