5-116
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Using The Work List
Section 5
7. Press
[START LOADER]
to begin sample processing.
8. The samples are processed in the order in which they were placed in the
racks. The Sample Loader automatically stops when all samples in the last
rack have been processed.
NOTE:
Additional entries may be made to the Work List while processing
takes place. To input additional entries to the Work List, press
[WORK LIST]
to display the
WORK LIST
screen. Type in the new
entries.
9. When all samples have been processed, check to see if all entries have been
deleted from the Work List. Entries are automatically deleted from the Work
List when a match occurs. Entries remaining in the Work List indicate a
successful match did not occur. The operator should follow laboratory
procedures to determine the cause of this discrepancy.
Using the Work List with Specimen IDs
When Work List entries contain Specimen IDs (non bar code), the operator
must
enter the rack and tube position of each sample into the <Rack and Tube #> field
of the appropriate Work List entry, otherwise no match can occur. It is not
necessary for the samples in the racks to be in the same order as the entries in the
Work List, but the position of each sample in the racks must match its Rack and
Tube # in the Work List.
Sample identification is made from the information entered in the
<Rack and Tube #>
6
th
field on the Work List.
If a match occurs (based on the Rack and Tube #), the sample results along with the
demographic data (including the Specimen ID) are displayed on the
RUN
screen
and sent to the Data Log.
NOTE:
The Rack and Tube # is not displayed on the
RUN
screen or sent to
the Data Log. The Rack and Tube # is used only for the purpose of
matching Work List entries with samples.
If a match does not occur, it indicates the operator typed an incorrect Rack and
Tube # into the Work List. The operator may either:
1. Correct the Rack and Tube # in the Work List and re-run the sample (this
avoids having to enter all the demographic data manually), OR
2. Go to the Data Log, find the entry by the Rack and Tube # assigned by the
instrument, and enter the remaining demographic information.
NOTE:
If the instrument cannot match a sample with a Work List entry
using Bar Code, Specimen ID, or Rack and Tube #, it identifies the
sample by its actual rack and tube position and assigns that number
to the <Specimen ID> field of that sample. The results along with
the assigned rack and tube number are sent to the Data Log
without
deleting any records in the Work List, since no match was made.
The sample is identified in the Data Log by the Rack and Tube #
assigned by the instrument.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...