CELL-DYN
®
3200 System Operator’s Manual
6-33
9140181H—October 2001
Section 6
Calibration Procedures
Enter Factor Method
Overview
The Enter Factor Calibration Method is used to enter a predetermined factor to
adjust calibration when a consistent bias exists between the CELL-DYN 3200
System and a comparison analyzer. This method can also be used when calibrating
the secondary mode of the system to the primary mode.
Either Calibrator or Fresh Whole Blood can be used for calibration.
Fresh Whole Blood: Because the same samples will be used to first obtain
reference values on a reference instrument then to calibrate in the Open and Closed
Modes, it is important to begin with a sufficient amount of sample (12 mL each is
recommended).
A set of worksheets is provided in
Manual Calibration Worksheets
at the end of
this section to assist in the calibration process.
NOTE:
Always complete the Pre-Calibration procedures before beginning any
calibration.
Enter Calibration Factor Screen
When the Enter Factor key is pressed in the main
CALIBRATION
menu, the
ENTER
CALIBRATION FACTO
R menu is displayed (refer to Figure 6.11). New calibration
factors are manually entered on this screen.
The following soft key labels are displayed:
RESTORE FACTORS
RESET ALL TO 1.000
RETURN
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...