10-92
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Troubleshooting and Diagnostics
Tables for Instrument Conditions and Messages
Section 10
The FAULT indicator light on the Analyzer status indicator panel is
illuminated in red
NOTE:
This type of fault is usually accompanied by a message in the status box
and/or on the bulletin line.
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The Analyzer has detected a fault
situation and has stopped operating.
1. From the first
DIAGNOSTICS MENU
screen, press
[FAULT REPORT]
a
copy of the report and
perform the
indicated
corrective action. When
the action is completed,
initialize the
Analyzer by
pressing the
[INITIALIZATION]
key on th
e
second
DIAGNOSTICS
MENU
screen.
2. If the fault report does not indicate a
message or action, document the
situation and initialize the Analyzer
by pressing the
[INITIALIZATION]
key
on the second
DIAGNOSTICS MENU
screen.
The message “Please check signal cable” appears on the Display Monitor
– – – - Probable Cause – – – –
– – – - Corrective Action – – –
1. The Display Monitor interface cable
is not securely connected.
1. Locate the VGA connector (25 pin)
on the Rear Panel of the Data
Module. Reseat the cable and secure
the connection properly.
2. Defective Data Module component.
2. Contact your Abbott Hematology
Customer Support Center.
Table 10.4 Non-Functional Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...