CELL-DYN
®
3200 System Operator’s Manual
10-73
9140181H—October 2001
Section 10
Troubleshooting and Diagnostics
Loader Warning 144: excessive cycling
/BL
NOTE:
The Sample Loader halts.
Twenty rack advances have occurred without a tube being sensed in the mix zone
or twenty-five rack advances have occurred after the last tube was processed.
– – – - Probable Cause – – –
– – – - Corrective Action –
1. The System is functioning as
intended.
1. No corrective action is necessary.
When you wish to run additional
samples in the Sample Loader, load
tubes in racks and press the keys in
the following order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESUME LOADER]
.
2. The rack advance mechanism is not
making contact with the racks.
2. Check for an obstruction that is
preventing the rack arms from
extending and holding the racks
against the loader wall so the racks
can engage with the rack advance
mechanism. If an obstruction is
found, remove it. Then, reset the
racks and reset the Sample Loader
by pressing the following keys in
order:
[CLEAR FAULT]
,
[START LOADER]
,
[RESET LOADER]
.
3. Both tube sensors are dirty.
3. Remove the Loader Cover. Using a
DYN-A-WIPE pad or lint-free
wipe moistened with lens cleaner or
deionized water, clean the tube
sensors. Then, reinstall the Loader
Cover, reset the racks, and reset the
Sample Loader.
NOTE:
For the location of the tube
sensor, refer to the figure
in
4. A failure of the sensor or related
electronics, or loader mechanisms
has occurred.
4. If the message appears repeatedly,
contact your Abbott Hematology
Customer Support Center.
Table 10.2 General Fault Conditions (Continued)
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...