6-50
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Calibration Procedures
Post Calibration Procedure
Section 6
2. Obtain SET-UP Disk#1 from the Disk Storage Container located on the back
of the instrument, and insert the disk into the instrument’s floppy disk drive
(located on the right-side panel of the instrument). Press and hold the
[CTRL]
and
[ALT]
keys, then press the
[DELETE]
key on the keyboard, this causes the
data station to boot the floppy disk.
3. When the DOS prompt displays:
Select (A) Abort; (S) Save; (R) Restore: [A, S, R]?
type “
s
” (Save). The following files are saved:
Nonvol
(NONVOL)
Calibration Log
(CALLOG)
Worklist Log
(WKLOG)
Error Log
(ERRORLOG)
Retic QC
(RQCLOG)
NOTE:
The (R) Restore option copies existing setup information from the
SET-UP Disk to the Data Station hard disk. Selecting “
r
” will
overwrite the current calibration and instrument setting files in the
CELL-DYN 3200 directory with previously-saved versions from
the archive on the floppy disk. In this case, the instrument may need
to be recalibrated and instrument settings may need to be
re-established by a field service representative.
This option is used when a hardware or software failure occurs and
should only be used at the direction of the Abbott Hematology
Customer Support Center.
NOTE:
In the message “Write Protect Error Writing Drive A” displays,
eject the SET-UP Disk. Unlock the disk by sliding the write-
protected tab on the disk’s corner. Reinsert the disk and press the
“
R
” (Retry) key on the keyboard.
4. When the following prompt displays, the backup is complete:
A:/
5. Remove SET-UP Disk #1 from the diskette drive and return it to the Disk
Storage Container located on the back of the instrument.
6. Turn the instrument power OFF then ON.
7. When the instrument is initialized, press the
[RUN]
key to prime the system.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...