5-100
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Operating Instructions
Daily Shutdown Procedure
Section 5
Peristaltic Pump Tubing
IMPORTANT:
By removing the tubing under the Sample Transfer Peristaltic
Pump, the life of the tubing is significantly prolonged. The tubing should be
removed if the instrument will be idle for more than 24 hours and reinserted prior
to operating the instrument.
Short Term Shutdown
If the shutdown time is expected to be 72 hours or less, follow the procedure below:
1. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
.
2. If the instrument is to be cleaned first, press
[AUTO CLEAN]
, otherwise go to
step 3.
3. Press
[DAILY SHUTDOWN]
.
4. Leave power ON.
5. Tubing should be removed from under the Peristaltic Pump.
Intermediate-term Shutdown
If the shutdown time is expected to exceed 72 hours (but less than 2 weeks), follow
the procedure below:
1. With the system still ON, perform any maintenance that is due.
2. In the
MAIN MENU
, press
[SPECIAL PROTOCOLS]
.
3. Press
[AUTO CLEAN]
to clean the instrument.
4. Press
[DAILY SHUTDOWN]
.
5. When the
'$,/<6+87'2:1
cycle is finished, turn power OFF in the
following order:
a. System (includes Analyzer, Data Module, and Sample Loader)
b. Display Monitor
c. Printer
NOTE:
In an emergency situation, turn OFF power in any order as quickly
as possible.
6. Remove the tubing from under the Sample Transfer Peristaltic Pump.
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...