11-18
CELL-DYN
®
3200 System Operator’s Manual
9140181H—October 2001
Quality Control
Quality Control Guide
Section 11
CELL-DYN 3200 Westgard
®
Rules
Overview
A control rule tests the control result against control limits to determine whether
the CELL-DYN 3200 shows acceptable accuracy and precision. The limits are
derived from the mean and standard deviation of control measurements obtained
when instrument performance is stable and acceptable. The most common rule
used in Hematology Quality Control is the mean ± 2SD limits. Ninety-five percent
of the control results should fall within the ± 2SD limits.
Quality control rules detect random or systematic error. Random error may be
defined as an increase in the SD (loss of precision). Systematic error may be
defined as a shift in the mean value (loss of accuracy). A multi-rule quality control
procedure combines several control rules to improve the detection of both types of
error.
Westgard recommended a multi-rule approach to evaluating quality control
results.
2
This approach has long been used in the chemistry laboratory
3
but is new
in the hematology laboratory. A set of modified Westgard
®
rules may be used to
monitor quality control results on the CELL-DYN 3200.
NOTE:
Do not use the values for mean range listed on the commercial control’s
assay sheet in conjunction with Westgard
®
Rules. Before using
Westgard
®
Rules with commercial controls, establish the SD for each
parameter on your instrument. Enter limits based on 2SDs. For Westgard
®
Rules, control limits are assumed to be 2SD.
Westgard
®
Rules
The modified Westgard
®
Rules (Westgard’s nomenclature is given in parentheses)
available on the CELL-DYN 3200 are:
Rule 1 (1
3s
):
Value outside 3SD
A control result exceeded the mean ± 3SD
Rule 2 (2
2s
):
Two consecutive values outside the same 2SD
Two consecutive results fell outside 2SD on the same side
of the mean
Rule 3 (R
4s
):
Two consecutive values outside opposite 2SD
One result was greater than 2SD above the mean and the
next result was greater than 2SD below the mean.
Consequently, the range between the results is greater
than 4SD.
Rule 4 (2 of 3
2s
):
Two of three consecutive values outside same 2SD
Two of the last three results fell outside 2SD on the same
side of the mean
Rule 5 (4
1s
):
Four consecutive values outside same 1SD
Four consecutive results fell outside 1SD on the same side
of the mean
Rule 6 (10x):
Ten consecutive values on the same side of mean
Ten consecutive results fell on the same side of the mean
Содержание CELL-DYN 3200
Страница 12: ...xii CELL DYN 3200 System Operator s Manual 9140181K July 2002 NOTES ...
Страница 52: ...1 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Use or Function Overview Section 1 NOTES ...
Страница 132: ...3 38 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Principles of Operation References Section 3 NOTES ...
Страница 154: ...5 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Overview Section 5 NOTES ...
Страница 234: ...5 82 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Operating Instructions Run Menu Section 5 NOTES ...
Страница 300: ...6 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures Overview Section 6 NOTES ...
Страница 356: ...6 58 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Calibration Procedures References Section 6 NOTES ...
Страница 362: ...8 2 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards Overview Section 8 NOTES ...
Страница 370: ...8 10 CELL DYN 3200 System Operator s Manual 9140181K July 2002 Hazards References Section 8 NOTES ...
Страница 372: ...9 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Service and Maintenance Overview Section 9 NOTES ...
Страница 532: ...11 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control Overview Section 11 NOTES ...
Страница 558: ...11 28 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Quality Control References Section 11 NOTES ...
Страница 560: ...12 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Printers Overview Section 12 NOTES ...
Страница 566: ...13 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Sample Loader Overview Section 13 NOTES ...
Страница 590: ...14 2 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Reticulocyte Package Overview Section 14 NOTES ...
Страница 662: ...A 4 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Overview Appendix A NOTES ...
Страница 666: ...A 8 CELL DYN 3200 System Operator s Manual 9140181H October 2001 Bar Codes Specifications Appendix A NOTES ...