Troubleshooting
Troubleshooting Guides
6-4
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Malfunction
Possible Cause
Action
No SHARP
Concentration
Disconnected Scat Det cable
Check cable connection.
No Zero Air Flow
Rear solenoid valve normally
closed
Check rear panel connection—
replace valve.
Clogged in-line filter or
internal filter
Replace filter.
Pump unplugged
Plug in pump.
Internal solenoid valve
normally closed
Check cable connection—replace
valve.
Check zero in bulkhead-tubing
connection
Kinked tube—insert tube.
Check pump exhaust
connection
Insert tube in “Y” fitting or pump.
Check internal “T” fitting for
plug or tubing not connected
Insert blocking port plug connect
tubing.
Ref Det Voltage
Low or high source current
Check source current level Adjust
to 62–67 mA
Check min and max values/LED
current 65mA
Neph Cal
Bad Zero
Re-zero
Neph Temp
High/Low
Check min and max values/
Re-calibrate
Neph RH
Low/High
Check min and max values/
Re-calibrate
Led Current
High/Low
Check min and max values/Verify
65mA ± 5mA