C-Link Protocol Commands
I/O Configuration
B-62
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Send:
analog vout range 2
Receive:
analog vout range 2 3
set analog vout range
channel range
This command sets analog voltage output
channel
to the range, where
channel
is between 1 and 6 inclusive, and
range
is set according to
Table B–
12
. The following example sets channel 2 to the 0-10 V range.
Send:
set analog vout range 2 3
Receive:
set analog vout range 2 3 ok
Table B–12.
Analog Voltage Output Range Values
Range
Output Range
1 0-1
V
2 0-100
mV
3 0-10
V
4 0-5
V
0 [cannot be set to this, but may report]
Undefined
dig in
This command reports the status of the digital inputs as a 4-digit
hexadecimal string with the most significant bit (MSB) being input 16.
Send:
dig
in
Receive:
dig in 0xffff
din
channel
This command reports the action assigned to the digital input
channel
and
the index number of the corresponding active state. The following example
reports input 1 to be assigned an index number 3 corresponding to action
of “PM mode” with the active state being high.
Send:
din
1
Receive:
din 1 3 PM MODE high
set din
channel index state
This command assigns digital input
channel
(1-16) to activate the action
indicated by
index
(1-12), when the input transitions to the designated
state
(high or low). Use the “list var din” command to obtain the list of
supported
index
values and corresponding actions. The following example
sets the digital input channel 1 to 3 on a low-to-high transition.
Send:
set din 1 3 high
Receive:
set din 1 3 high ok
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......