Operation
Volumetric Conditions
3-34
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
●
In the Main Menu, choose Instrument Controls > Datalogging
Settings > Configure Datalogging >
Memory Allocation %
.
●
●
●
●
●
The Data Treatment screen is used to select the data type for the selected
record type: whether the data should be averaged over the interval, the
minimum or maximum measured during the interval, or the current value
(last value measured). Data treatment doesn’t apply to all data, just to the
concentration measurement. All other data points log the current value at
the end of the interval.
Note
This feature is found in all
i
Series instruments, but it is
recommended that the data type be set to ONLY the current value (cur), as
the datalogging averaging is done in addition to the normal concentration
averaging.
▲
●
In the Main Menu, choose Instrument Controls > Datalogging
Settings > Configure Datalogging >
Data Treatment
.
SET LREC DATA TYPE:
CURRENTLY: AVG
SET TO: CUR ?
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Flag Status Data screen is used to set the flag status data to either ON
or OFF for the selected record type.
●
In the Main Menu, choose Instrument Controls > Datalogging
Settings > Configure Datalogging >
Flag Status Data
.
Data Treatment
Flag Status Data
SET PERCENT LRECS:
CURRENTLY: 50%
SET TO: 30% ?
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
SET PERCENT LRECS:
CURRENTLY: 50%
SET TO: 30% ?
THIS WILL ERASE ALL LOGS!
x
PRESS
TO CONFIRM CHANGE
RANGE AVG DIAGS ALARM
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......