Servicing
Delrin Nut Replacement
7-50
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
2.
Turn the Heater OFF: From the Main Menu, choose Instrument
Controls > Set Heater >
Control
. Use
until the value reads
OFF, and press
.
3.
Loosen the two hex screws from the collar of the heater adapter cover.
4.
Loosen the two captive screws of the heater adaptor cover and slide up.
5.
Loosen the bottom 5/8-inch union nut and remove the tube adapter.
6.
Slide off the heater tube adapter down.
7.
Slide off ferrite bead upward off the heater tube adapter.
8.
Slide off delrin nut upward off the heater tube adapter.
9.
Replace the delrin nut by following the previous steps in reverse.
Figure 7–29.
Replacing the Delrin Nut
Heater Tube Adapter
Union Nut
Ferrite Bead
Hex Screws (2)
Delrin Nut
Captive Screws (2)