C-Link Protocol Commands
I/O Configuration
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
B-65
17
frol
19
beta
28
braw
29
araw
30
avgpm
32
mass
34
bzero
45
exfg
50
bref
52
cflg
53
aflg
relay stat
This command reports the current relay logic as normally “open” or
normally “closed,” if all the relays are set to same state, that is all open or all
closed. The following example shows that the status of all the relays’ logic is
set to normally “open”.
Send:
relay
stat
Receive:
relay stat open
Note
If individual relays have been assigned different logic, then the
response would be a 4-digit hexadecimal string with the least significant
byte (LSB) being relay no 1.
▲
For example:
Receive:
relay stat 0x0001
(indicates relay no 1 is set to normally open logic, all others are normally
closed)
Receive:
relay stat 0x0005
(indicates relay no 1 and 3 are set to be normally open logic, all others are
normally closed)
set relay open
set relay open
value
set relay closed
set relay closed
value
These commands set the relay logic to normally open or closed for relay
number
value
, where
value
is the relay between 1 and 16. The following
example sets the relay no 1 logic to normally open.
Note
If the command is sent without an appended relay number, then all
the relays are assigned the set logic of normally open/closed.
▲
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......