Servicing
Removing the SHARP Optics Cover
7-44
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
The SHARP optics cover can be removed from the SHARP optics. Refer to
the following steps when a procedure requires access to the SHARP optics
assembly (
Figure 7–24
).
Figure 7–24.
Removing the SHARP Optics Cover
Equipment Required:
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF and unplug the power cord.
2.
If the instrument is mounted on a rack, remove it from the rack.
3.
Unfasten the two screws.
4.
Lift cover in an upward direction.
5.
Replace the SHARP optics cover by following the previous steps in
reverse.
Removing the SHARP
Optics Cover
Remove Screw
Remove Screw