Operation
Diagnostics Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-59
PROGRAM VERSIONS:
PRODUCT: MODEL 5030i
VERSION: 00.00.32.073
FIRMWARE: 10.22.95
RANGE AVG DIAGS ALARM
The Voltages menu displays the current diagnostic voltage readings. This
screen enables the power supply to be quickly read for low or fluctuating
voltages without having to use a voltage meter. The I/O board item is only
displayed if the I/O expansion board option is installed.
●
In the Main Menu, choose Diagnostics >
Voltages
.
VOLTAGES:
>MOTHERBOARD
INTERFACE BOARD
I/O BOARD
DETECTOR BOARD
RANGE AVG DIAGS ALARM
The Motherboard screen (read only) is used to display the voltage readings
on the motherboard.
●
In the Main Menu, choose Diagnostics > Voltages >
Motherboard
.
MOTHERBOARD VOLTAGES:
3.3 SUPPLY 3.3 V
5.0 SUPPLY 5.0 V
15.0 SUPPLY 15.0 V
24.0 SUPPLY 24.1 V
-3.3 SUPPLY -3.3 V
RANGE AVG DIAGS ALARM
The Interface Board screen (read only) is used to display the voltage
readings on the measurement interface board.
●
In the Main Menu, choose Diagnostics > Voltages >
Interface
Board
.
Voltages
Motherboard Voltages
Interface Board Voltages