Installation
Data Content
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
2-23
set to 60 minute logging with 30% memory and Srecs are 5-
minute logging with 70% memory allocated. If there are any
changes, be sure to Commit Content.
b.
From the Main Menu, choose Service. Under Service:
Note
If Service Mode is not displayed, refer to “
Accessing the Service
Mode
” on page
7-4
, then return to the beginning of this step.
▲
i.
Check/Calibrate Ambient Temperature, Nephelometer
Temperature, and Flow Temperature
ii.
Check/Calibrate Nephelometer Ambient RH
iii.
Check Calibrate Barometric Pressure Span
iv.
Check/Calibrate Flow Calibration
v.
Perform a Beta Auto Detector Calibration (30 minutes)
vi.
Perform a Mass Foil Calibration (factory calibrated)
vii.
Mass Coefficients are set at the factory.
viii.
Perform a Nephelometer Zero Calibration (5 minutes)
Please use
i
Port to access data download from the instrument.
Use the following procedure when shutting down the instrument.
1.
Turn the power OFF.
2.
Cover instrument ports when not in use.
DO NOT
●
Do not pressurize the monitor.
●
Do not allow liquids to enter the sample port.
●
Do not expose the instrument to excessive vibration or magnetic
interference.
●
Do not remove/adjust the beta attenuation bench of the 5030
i
.
●
Do not remove the C-14 radioactive source at any time.
Data Content
Shutdown
Important Tips
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......