C-Link Protocol Commands
Alarms
B-20
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Send:
set alarm flow pres max 50
Receive:
set alarm flow pres max 50 ok
alarm flow temp min
alarm flow temp max
This command reports the current flow temperature alarm maximum value
setting. The following example reports that the flow temperature alarm
maximum value is 20.0 °C.
Send:
alarm flow temp max
Receive:
alarm flow temp max 20.0 degC
set alarm flow temp min
value
set alarm flow temp max
value
This command sets the flow temperature alarm maximum value to
value
,
where
value
is a floating-point number representing flow temperature alarm
limits in °C. The following example sets the flow temperature alarm
maximum value to 70 °C.
Send:
set alarm flow temp max 70
Receive:
set alarm flow temp max 70 ok
alarm ib status
This command reports the status of the interface board as FAIL or OK.
The following example reports that the interface board status is OK.
Send:
alarm ib status
Receive:
alarm ib status ok
alarm io status
This command reports the status of the I/O expansion board as FAIL or
OK. The following example reports that the I/O expansion board status is
OK.
Send:
alarm io status
Receive:
alarm io status ok
alarm mb status
This command reports the status of the motherboard as FAIL or OK. The
following example reports that the motherboard status is OK.
Send:
alarm mb status
Receive:
alarm mb status ok
alarm nb status
This command reports the status of the nephelometer interface board as
FAIL or OK. The following example reports that the nephelometer
interface board status is OK.
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......