Servicing
Measurement Interface Board Replacement
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-31
Use the following procedure to replace the measurement interface board
(
Figure 7–15
).
Equipment Required:
Measurement interface board
Philips screwdriver
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Lower the partition panel, then proceed to the next step below. Refer to
“
Removing the Measurement Case Assembly and Lowering the
Partition Panel
” in this chapter.
2.
Unplug all connectors from the measurement interface board. Note the
locations of the connectors to facilitate re-connection.
3.
Unscrew the two screws at the top of the measurement interface board.
Pop the measurement interface board off from the two bottom
mounting studs and remove the board.
4.
To install the measurement interface board, follow the previous steps in
reverse.
5.
Re-install the measurement case assembly.
6.
Calibrate the instrument. Refer to the “
Calibration
” chapter in this
manual.
Measurement
Interface Board
Replacement
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......