Operation
Volumetric Conditions
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-37
STOP BITS:
CURRENTLY: 1
SET TO: 2 ?
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
The RS-232/RS-485 Selection screen allows the user to choose between the
RS-232 or RS-485 specification for serial communication.
Equipment Damage
Disconnect the serial cable before changing the RS-
232 and RS-485 selection to prevent damage to any equipment currently
connected to the analyzer.
▲
●
In the Main Menu, choose Instrument Controls > Communication
Settings > Serial Settings >
RS-232/485 Selection
.
●
●
●
●
●
The Instrument ID screen allows the operator to edit the instrument ID
number. The ID is used to identify the instrument when using the C-Link
or MODBUS protocols to control the instrument or collect data. It may be
necessary to edit the ID number if two or more instruments of the same
model are connected to one computer. Valid instrument ID numbers are
from 0 to 127. The Model 5030
i
has a default instrument ID of 14. For
more information about the instrument ID, see Appendix B “
C-Link
Protocol
” or Appendix C “
MODBUS Protocol
”.
●
In the Main Menu, choose Instrument Controls > Communication
Settings >
Instrument ID
.
RS-232/RS-485 Selection
Instrument ID
RS-232/RS-485 SELECT:
XXXXXX
** WARNING **
XXXXXXX
DISCONNECT THE SERIAL
X X
CABLES BEFORE CHANGING
X
THE SELECTION!
XXXXXX
TO CONTINUE
RANGE AVG DIAGS ALARM
RS-232/RS-485 SELECT:
CURRENTLY: RS-232
SET TO: RS-485 ?
MAKE SURE THAT THE CABLE
XX
IS OFF: PRESS
TO CONFIRM
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......