Operation
Alarms Menu
3-68
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
DETECTOR ALARMS:
BOARD STATUS OK
>ALPHA OK
BETA OK
RANGE AVG DIAGS ALARM
The Alpha Alarms screen allows the user to view the minimum and
maximum alarm parameters. Acceptable alarm limits range from 0 to 100.
If the alpha count reading goes beyond the maximum limit, an alarm is
activated and the alarm (bell) icon appears in the status bar on the Run
screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Detector Alarms >
Alpha
.
ALPHA COUNT:
ACTUAL 0
MIN 0
MAX 100
RANGE AVG DIAGS ALARM
The Beta Alarms screen allows the user to view the minimum and
maximum alarm parameters. Acceptable alarm limits range from 5000 to
20000. If the beta count reading goes beyond the maximum limit, an alarm
is activated and the alarm (bell) icon appears in the status bar on the Run
screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Detector Alarms >
Beta
.
BETA COUNT:
ACTUAL 0
MIN 5000
MAX 20000
RANGE AVG DIAGS ALARM
The Nephelometer Alarms submenu allows the user to view and set the
nephelometer alarm parameters. Board status indicates that the power
supplies are working and connections are successful. The nephelometer
Min and Max Alpha
Alarms
Min and Max Beta
Alarms
Nephelometer
Alarms
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......