C-Link Protocol Commands
Measurement Configuration
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
B-47
filter time
This command reports the next date and time of the filter tape change. The
following example reports that the date and time is January 1, 2008, 6:12
pm.
Send:
filter
time
Receive:
filter time 01Jan08 18:12
set filter time
dd-mm-yyyy hh:mm:ss
dd
= day
mm
= month
yyyy
= year
hh
= hours
mm
= minutes
ss
= seconds
This command sets the next filter tape change. The following example sets
the filter time to October 1, 2008, 11:32 pm.
Send:
set filter time 10-01-2008 23:32
Receive:
set filter time 10-01-2008 23:32 ok
flow target
This command reports the current flow. The following example reports
that the flow is 16.0 LPM.
Send:
flow
target
Receive: flow
target
16.00
set flow target
value
This command sets the flow to
value
, where
value
is a floating-point
number representation of flow between 1 to 20 LPM. The following
example sets the flow to 16.67 LPM.
Send:
set flow target 16.67
Receive:
set flow target 16.67 ok
ht control
This command reports the current status of the heater (off, RH threshold,
or temp threshold). The following example reports that the heater is RH
threshold.
Send:
ht
control
Receive:
ht control RH
set ht control
selection
selection
= | off | rh | temp |
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......