Servicing
Pressure Board Replacement
7-18
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
4.
Install the new pump by following the previous steps in reverse.
5.
Perform a leak test as described in the “
Preventive Maintenance
”
chapter.
Figure 7–8.
Replacing the Pump
Use the following procedure to replace the pressure board (
Figure 7–9
).
Equipment Required:
Pressure board
Philips screwdriver, #2
Note
If your pressure board is on the base plate, order replacement retrofit
kit and follow instructions included with kit.
▲
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover.
2.
Disconnect plumbing from the pressure board assembly (
Figure 7–10
).
Note the plumbing connections to facilitate reconnection.
Pressure Board
Replacement
Connect to Rear Panel Elec Pump
Connect SHARP Valve
Port 1
PUMP
Connect to Rear
Panel “VACUUM”