Operation
Volumetric Conditions
3-54
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
ANALOG INPUT UNITS:
CURRENTLY: V
V
N1
ABCDEFGHIJKLMN BKSP
OPQRSTUVWXYZ PAGE
0123456789 ./- SAVE
RANGE AVG DIAGS ALARM
The Decimal Places screen allows the user to select how many digits are
displayed to the right of the decimal point, from 0 to 6, with a default of 2.
●
In the Main Menu, choose Instrument Controls > I/O Configuration
> Analog Input Config > select Channel >
Decimal Places
.
DECIMAL PLACES:
CURRENTLY: 2
SET TO: 3 ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Number of Table Points screen allows the user to select how many
points are used in the analog input conversion table for the selected
channel. The instrument uses linear interpolation between the points in
this table to determine what the reading value is based on the analog input
voltage. Each point in the table consists of an analog input voltage value
(0–10.5 V) and a corresponding reading value. Only two points are
necessary for linear inputs, however a larger number of points may be used
to approximate non-linear inputs. The points range from 2 to 10, with a
default of 2.
●
In the Main Menu, choose Instrument Controls > I/O Configuration
> Analog Input Config > select Channel >
Table Points
.
NUMBER OF TABLE POINTS:
CURRENTLY: 2
SET TO: 10 ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
Decimal Places
Number of Table Points
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......