Operation
Alarms Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-67
INSTRUMENT ALARMS:
>FILTER TAPE COUNTER OK
FILTER TAPE CHANGE OK
BENCH OK
HEATER POWER OK
MOTHERBOARD STATUS OK
RANGE AVG DIAGS ALARM
INTERFACE STATUS OK
I/O EXP STATUS OK
The Filter Tape Counter screen displays the current filter tape counter
value and sets the maximum alarm limit. If the filter tape counter reading
goes beyond the maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Instrument Alarms >
Filter
Tape Counter
.
FILTER TAPE COUNTER:
ACTUAL 0
>MAX 650
RANGE AVG DIAGS ALARM
The Maximum Filter Tape Counter screen is used to change the maximum
filter tape counter alarm limit (480 = 90% usage, 10% remaining).
●
In the Main Menu, choose Alarms > Instrument Alarms > Filter Tape
Counter >
Max
.
FILTER TAPE COUNTER:
ACTUAL MAX: 650
SET MAX TO:
xxxx
48
0
?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Detector Alarms submenu allows the user to view a set of alarms for
detector boards. Board status indicates that the power supplies are working
and connections are successful. There are no setting screens for this alarm.
●
In the Main Menu, choose Alarms >
Detector Alarms
.
Filter Tape Counter
Max Filter Tape Counter
Detector Alarms
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......