C-Link Protocol Commands
Communications Configuration
B-60
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
set power up mode
value
This command is used to configure the instrument to power up in the
local/unlocked mode (
value
= 0
)
or the remote/locked mode (
value
= 1),
according to
Table B–10
.
If the instrument is set to power up in the local/remote mode, the keypad
can be used to make changes via the front panel. If the instrument is set to
power up in the remote/locked mode, changes can not be made from the
front panel. The following example sets the instrument to power up in
remote/locked mode.
Send:
set
power
up
mode
1
Receive:
set power up mode 1 ok
Table B–10.
Power Up Mode Values
Value
Power up Mode
0 Local/Unlocked
Mode
(default)
1 Remote/Locked
Mode
program no
This command reports the analyzer’s model information and program
version number.
Send:
program
no
Receive:
program no iSeries 5030i 00.05.37.093
tz
This command reports the “tz” timezone string for the NTP server. See
“Network Time Protocol Server” in the “Communications Settings”
section of the “Operation” chapter for more information.
Send:
tz
Receive: tz
EST+5EDT
set tz
string
This command sets the timezone
string
for the instrument for use with the
NTP server, where
string
is a standard timezone string. Common strings are
listed in the “Timezone” screen description in Chapter 3.
Send:
set tz EST+5EDT
Receive:
set tz EST+5EDT ok
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......