Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
D-6
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
The Gesytec serial number defaults to zero. To set the Gesytec serial
number select Main Menu > Instrument Controls > Communication
Settings >
Gesytec Serial No
.
Example of response to DA command from an instrument with Gesytec
serial number set to 000.
The Gesytec serial number is
bold
in the
example.
Gesytec Protocol with transmission of three concentrations (Instrument ID
is 1, Operation Status is 03, Error Status is 04):
Data Query String:
<STX>DA<CR>
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>
000
0000000 <SP>002 <SP>
Address First Concentration(E-format)=25.78 1
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681 2 Third Concentration=11.75
000
0000000<SP><CR>
Example of response to DA command from an instrument with Gesytec
serial number set to 123.
The Gesytec serial number is
bold
in the
example.
Gesytec Protocol with transmission of three concentrations (Instrument ID
is 1, Operation Status is 03, Error Status is 04):
Data Query String:
<STX>DA<CR>
Reply String:
<STX>MD03<SP>001<SP>+2578+01<SP>03 <SP>04<SP>
123
0000000 <SP>002 <SP>
Address First Concentration(E-format)=25.78 1
+5681+00<SP>03<SP>04<SP>0000000000<SP>003<SP>+1175+01<SP>03<SP>04<SP
Second Concentration = 5.681 2 Third Concentration=11.75
123
0000000<SP><CR>
The attached concentrations are in the selected gas units. The
measurements that are attached to the response, if not valid in a particular
mode, will report a value of 0.0.
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......