Servicing
Delrin Nut Replacement
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-49
Figure 7–28.
Replacing the SHARP Optics Board
Use the following procedure to replace the Delrin nut (
Figure 7–29
).
Equipment Required:
Delrin nut
Philips screwdriver
Set of standard hex drives
Adjustable wrench
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the Pump Flow OFF: From the Main Menu, choose Instrument
Controls > Set Flow/Pump >
Pump
. Press
to toggle and set the
pump flow to OFF.
Delrin Nut
Replacement
Loosen screws
Loosen screws
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......