Operation
Volumetric Conditions
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-55
The Table Point submenu allows the user to set up an individual table
point.
●
In the Main Menu, choose Instrument Controls > I/O Configuration
> Analog Input Config > select Channel >
Point 1–10
.
TABLE POINT 01 CONFIG:
>VOLTS 0.00
USER VALUE 0.00
RANGE AVG DIAGS ALARM
The Volts screen allows the user to set the input voltage for the selected
table point in the conversion table, from 0.00 to 10.50. The default table is
a two-point table with point 1: 0.00 V = 000.0 U and point 2: 10.00 V =
10.0 U, where U is the previously entered unit of measure.
●
In the Main Menu, choose Instrument Controls > I/O Configuration
> Analog Input Config > select Channel > select Point >
Volts
.
TABLE POINT 01 VOLTS:
CURRENTLY: 0.00
SET TO: 00.0
0
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The User Value screen allows the user to set the output value for the
corresponding input voltage for the selected table point in the conversion
table, from -9999999 to 99999999. The default table is a two-point table
with point 1: 0.00 V = 000.0 U and point 2: 10.00 V = 10.0 U, where U is
the previously entered unit of measure.
●
In the Main Menu, choose Instrument Controls > I/O Configuration
> Analog Input Config > select Channel > select Point >
User Value
.
Table Points
Volts
User Value
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......