C-Link Protocol Commands
Alarms
B-14
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Send:
pm
24avg
Receive:
pm 24avg 10.000E+00 ug/m3
sharp
This command reports the 1-minute running average SHARP
concentration. The following example reports the 1-minute running
average value.
Send:
sharp
Receive: sharp
1.187E+01
μ
g/m3
sharp 24avg
This command reports the last calculated 24-hour average SHARP
concentration (which defaults to 9,999 μg/m
3
if 24 hours has not passed).
Send:
sharp
24avg
Receive: sharp
24avg
9.999E+03
μ
g/m3
alarm alpha count min
alarm alpha count max
These commands report the current alpha count alarm minimum and
maximum value settings. The following example reports that the alpha
count alarm minimum value is 200.
Send:
alarm alpha count min
Receive:
alarm alpha count min 200
alarm amb rh min
alarm amb rh max
These commands report the current ambient relative humidity alarm
minimum and maximum value settings. The following example reports the
ambient relative humidity alarm minimum value is 35.0%.
Send:
alarm amb rh min
Receive:
alarm amb rh min 35.0 %
set alarm amb rh min
value
set alarm amb rh max
value
These commands set the ambient relative humidity alarm minimum and
maximum values to
value
, where
value
is a floating-point number
representing ambient relative humidity alarm limits in percent. The
following example sets the ambient relative humidity alarm maximum
value to 55%.
Send:
set alarm amb rh max 55
Receive:
set alarm amb rh max 55 ok
Alarms
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......