Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
D-8
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
See
Table D–1
for operating status and
Table D–2
for error status for the
Model 5030
i
.
Table D–1.
Operating Status for Model 5030
i
D7 D6 D5 D4 D3 D2 D1 D0
Bit
8
7
6
5
4
3
2
1
Hex-value
80 40 20 10 08 04 02 01
MSB
LSB
Operating status:
Service
Mode
(On)
0 0 0 0
0 0 0 1
No
Password
Protect
0 0 0 0
0 0 1 0
Pump
(Off)
0 0 0 0
0 1 0 0
Heater
(Off)
0 0 0 0
1 0 0 0
Filter
Change
0 0 0 1
0 0 0 0
Nephelometer
Zero
0 0 1 0
0 0 0 0
Nephelometer
Standby
0 1 0 0
0 0 0 0
Not
Used
1 0 0 0
0 0 0 0
Table D–2.
Error Status for Model 5030
i
D7 D6 D5 D4 D3 D2 D1 D0
Bit
8
7
6
5
4
3
2
1
Hex-value
80 40 20 10 08 04 02 01
MSB
LSB
Error status:
Filter Tape Cntr Alarm
0
0
0
0
0
0
0
1
Filter
Change
0 0 0 0 0 0 1 0
Beta
Cnts
Alarm
0 0 0 0 0 1 0 0
Neph
Brd
Alarm
0 0 0 0 1 0 0 0
Detector Board Status Alarm
0
0
0
1
0
0
0
0
Motherboard
Status
Alarm
0 0 1 0 0 0 0 0
Flow
Alarm
0 1 0 0 0 0 0 0
Nephelometer Reference Voltage
Alarm
1 0 0 0 0 0 0 0
Operating and Error
Status
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......