Operation
Alarms Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-75
The Minimum Flow Temperature alarm limit screen is used to change the
minimum flow temperature alarm limit. The minimum and maximum
flow temperature screens function the same way.
●
In the Main Menu, choose Alarms > RH/Temperature Alarms >
Flow Temp >
Min
or
Max
.
FLOW TEMPERATURE:
ACTUAL MIN: 0.0
o
C
SET MIN TO: 5.5
o
C?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Board Temperature screen displays the current board temperature
value and sets the minimum and maximum alarm limits. Acceptable alarm
limits range from 5 to 60 °C. If the board temperature reading goes beyond
the minimum or maximum limit, an alarm is activated and the alarm (bell)
icon appears in the status bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > RH/Termperature Alarms >
Board Temp
.
BOARD TEMP:
ACTUAL 26.1
o
C
>MIN 0.0
o
C
MAX 60.0
o
C
RANGE AVG DIAGS ALARM
The Minimum Board Temperature alarm limit screen is used to change the
minimum board temperature alarm limit. The minimum and maximum
board temperature screens function the same way.
●
In the Main Menu, choose Alarms > RH/Temperature Alarms >
Board Temp >
Min
or
Max
.
Min and Max Flow
Temperature Limits
Board Temperature
Min and Max Board
Temperature Limits
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......