Calibration
Calibration Procedure
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
4-7
●
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Vac/Flow >
Flow Pres Span
.
CAL FLOW PRESSURE SPAN:
FLOW: 21.0 mmHg
SET TO: 20.
0
mmHg?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Be sure to save the entry and compare the values once more. Repeat as
necessary to within 2 mmHg.
The preferred method of flow rate calibration is the Auto Flow Calibration.
Place a NIST-traceable volumetric flow meter on top of the instrument
after complete installation (reassembled heater tube, sample tubes, and
inlets). Allow approximately 1-minute for the flow to stabilize. Once stable,
measure and average three individual readings from the reference. This
average observed flow rate from your reference meter should now be
entered in the screen below:
FLOW AUTO CALIBRATION:
FLOW: 16.67 LPM
SET TO: 17.2
5
LPM?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Be sure to save the entry and compare the values once more. Repeat as
necessary to /- 2 %.
Perform the “
Leak Test
” procedure on page
5-6
to verify no leaks.
The Model 5030
i
SHARP is calibrated for mass in the factory using a series
of null and span foils. The mass transfer standard foil sets are available from
Thermo Fisher Scientific, and should be used as part of a QA Program for
performing a QC check on the mass measurements. This mass foil
calibration procedure can be used for QC checks, auditing, and calibration.
Flow Calibration
Auto Flow Calibration
Mass Calibration
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......