C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
B-17
alarm conc avg pm min
alarm conc avg pm max
These commands report the current 24-hour average PM concentration
alarm minimum and maximum values setting. The following example
reports that the average PM concentration minimum is 100 μg/m
3
.
Send:
alarm conc avg pm min
Receive:
alarm conc avg pm min 1.000E+02
μ
g/m3
set alarm conc avg pm min
value
set alarm conc avg pm max
value
These commands set the 24-hour average PM concentration alarm
minimum and maximum values to
value
, where
value
is a floating-point
representation of the concentration alarm limits. Values must be in the
units that are currently set for use. The following example sets the average
PM concentration alarm maximum value to 10000 mg/m
3
.
Send:
set alarm conc avg pm max 10000
Receive:
set alarm conc avg pm max 10000 ok
alarm conc avg sharp min
alarm conc avg sharp max
These commands report the current 24-hour average SHARP
concentration alarm minimum and maximum values setting. The following
example reports that the average SHARP concentration minimum is 0.0
μg/m
3
.
Send:
alarm conc avg sharp min
Receive:
alarm conc avg sharp min 0.000E+00
μ
g/m3
set alarm conc avg sharp min
value
set alarm conc avg sharp max
value
These commands set the 24-hour average SHARP concentration alarm
minimum and maximum values to
value
, where
value
is a floating-point
representation of the concentration alarm limits. Values must be in the
units that are currently set for use. The following example sets the average
SHARP concentration alarm maximum value to 10,000 μg/m
3
.
Send:
set alarm conc avg sharp max 10000
Receive:
set alarm conc avg sharp max 10000 ok
alarm conc inst neph min
alarm conc inst neph max
These commands report the current 24-hour instant nephelometer
concentration alarm minimum and maximum values setting. The following
example reports that the instant nephelometer concentration minimum is 0
μg/m
3
.
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......