C-Link Protocol Commands
Calibration
B-42
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
neph restore bkg
This command will restore the background values to the previous
background values. The following example sets this command.
Send:
set neph restore bkg
Receive:
set neph restore bkg ok
neph rh offset
This command reports the nephelometer relative humidity sensor offset set
at the factory. The following example reports that the nephelometer relative
humidity sensor offset is 7.3%.
Send:
neph rh offset
Receive:
neph rh offset 7.3 %
set neph rh offset
value
This command sets the nephelometer relative humidity sensor offset to
user-defined values to
value
, where
value
is a floating-point representation
of the offset. The following example sets the nephelometer relative
humidity sensor offset to 3%.
Send:
set neph rh offset 3
Receive:
set neph rh offset 3 ok
neph src level
This command reports the nephelometer source level set at the factory. The
following example reports that the nephelometer source level is 44%.
Send:
neph src level
Receive:
neph src level 44 %
set neph src level
value
This command sets the nephelometer source level to user-defined values to
value
, where
value
is a floating-point representation of the level. The
following example sets the nephelometer source level to 50%.
Send:
set neph src level 50
Receive:
set neph src level 50 ok
neph temp offset
This command reports the nephelometer temperature offset set at the
factory. The following example reports that the nephelometer temperature
offset is 2.6 °C.
Send:
neph
temp
offset
Receive:
neph temp offset 2.6 degC
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......