C-Link Protocol Commands
Calibration
B-36
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
set baro mass coef
value
This command sets the barometric mass coefficient to user-defined values
to
value
, where
value
is a floating-point representation of the coefficient.
The following example sets the barometric mass coefficient to 0.000630.
Send:
set baro mass coef 0.000630
Receive:
set baro mass coef 0.000630 ok
baro span
This command reports the current barometric pressure span recorded at the
time of calibration. The following example reports that the barometric span
is 1.0.
Send:
baro
span
Receive: baro
span
1.0000
set baro span
value
This command requires to manually specifying the span
value
for the
barometric pressure span, where
value
is a floating-point representation of
barometric pressure span. The following example successfully sets the
barometric span to 1.023.
Send:
set baro span 1.023
Receive:
set baro span 1.023 ok
beta th
This command reports the current reference beta threshold for the
detector. The following example reports that the beta threshold is 440.
Send:
beta
th
Receive:
beta th 440
set beta th
value
This command sets the reference beta threshold for the detector to
value
,
where
value
is a floating-point representation of the beta threshold. The
following example sets the beta threshold to 500.
Send:
set beta th 500
Receive:
set beta th 500 ok
bkg pm
This command reports the current PM backgrounds. The following
example reports that the PM background is 4 μg/m
3
.
Send:
bkg
pm
Receive:
bkg pm 4.000E+00
g/m3
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......