Operation
Service Menu
3-86
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
RH/TEMP CALIBRATION:
>AMBIENT TEMP 4.2
o
C
FLOW TEMP 0.8
o
C
AMBIENT RH 0.0
%
RANGE AVG DIAGS ALARM
The Ambient Temperature screen allows the user to view and calibrate
ambient temperature. Adjust the zero offset until the ambient temperature
agrees with your treaceable standard.
●
In the Main Menu, choose Service > RH/Temp Calibration >
Ambient Temp
.
CALIBRATE AMBIENT TEMP:
TEMPERATURE: 29.9
o
C
OFFSET: 4.2
o
C
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Flow Temperature screen allows the user to view and calibrate flow
temperature. Be sure to have the cover removed and equilibrated to room
temperature.
●
In the Main Menu, choose Service > RH/Temp Calibration >
Flow
Temp
.
CALIBRATE FLOW TEMP:
TEMPERATURE: 28.0
o
C
OFFSET: 0.8
o
C
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Ambient Relative Humidity screen allows the user to view and
calibrate relative humidity. Adjust the zero offset until the RH agrees with
your treaceable standard.
●
In the Main Menu, choose Service > RH/Temp Calibration >
Ambient RH
.
Ambient Temperature
Flow Temperature
Ambient Relative
Humidity