C-Link Protocol Commands
Alarms
B-18
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Send:
alarm conc inst neph min
Receive:
alarm conc inst neph min 0.000E+00
μ
g/m3
set alarm conc inst neph min
value
set alarm conc inst neph max
value
These commands set the 24-hour instant nephelometer concentration
alarm minimum and maximum values to
value
, where
value
is a floating-
point representation of the concentration alarm limits. Values must be in
the units that are currently set for use. The following example sets the
instant nephelometer concentration alarm maximum value to 10000
μg/m
3
.
Send:
set alarm conc inst neph max 10000
Receive:
set alarm conc inst neph max 10000 ok
alarm conc inst pm min
alarm conc inst pm max
These commands report the current 24-hour instant PM concentration
alarm minimum and maximum values setting. The following example
reports that the instant PM concentration minimum is 100 μg/m
3
.
Send:
alarm conc inst pm min
Receive:
alarm conc inst pm min 1.000E+02
μ
g/m3
set alarm conc inst pm min
value
set alarm conc inst pm max
value
These commands set the 24-hour instant PM concentration alarm
minimum and maximum values to
value
, where
value
is a floating-point
representation of the concentration alarm limits. Values must be in the
units that are currently set for use. The following example sets the instant
PM concentration alarm maximum value to 10000 μg/m
3
.
Send:
set alarm conc inst pm max 10000
Receive:
set alarm conc inst pm max 10000 ok
alarm conc inst sharp min
alarm conc inst sharp max
These commands report the current 24-hour instant SHARP concentration
alarm minimum and maximum values setting. The following example
reports that the instant SHARP concentration minimum is 0 μg/m
3
.
Send:
alarm conc inst sharp min
Receive:
alarm conc inst sharp min 0.000E+00
μ
g/m3
set alarm conc inst sharp min
value
set alarm conc inst sharp max
value
These commands set the 24-hour instant SHARP concentration alarm
minimum and maximum values to
value
, where
value
is a floating-point
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......