Gesytec (Bayern-Hessen) Protocol
Gesytec Commands
D-4
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
If the received command does not satisfy the above formats, or if the
<address> does not match the Instrument Address, the command is
ignored.
This is a sample command to switch the instrument to initiate a filter
change, instrument address 14:
<STX>ST014<SP>N<CR>
This command initiates a data transfer from the instrument. The
instrument responds with measurement data, which depends on the range
mode and is listed in “
Measurements reported in response to DA
command
” below.
The command structure for a data query command is as follows:
<STX>DA<address><ETX><BCC>
The <address> is optional, which means it can be left out completely. The
<address>, if present, must match the Instrument Address. Additional space
can be present after the <address>.
If the <address> is left out, then no space is allowed in the query string.
A command with no address is also a valid command.
The following are the different acceptable formats of the DA command
with Instrument Address 14:
<STX>DA<CR>
<STX>DA014<CR>
<STX>DA<SP><14<ETX><BCC>
<STX>DA<ETX><BCC>
The data query string is valid and will be answered with data transmission
only if the command starts with <STX>, which is followed by the
characters DA, and the <address> (if present) matches the Instrument
Address, and the command is terminated with either <CR> with no
checksum or <ETX>, followed by the correct checksum <BCC>.
Data Sampling/Data
Query Command
(DA)
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......