Servicing
Removing the Measurement Case Assembly and Lowering the Partition Panel
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
7-13
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling any internal component.
▲
1.
Turn the instrument OFF and unplug the power cord.
2.
If the instrument is mounted in a rack, remove it from the rack.
3.
Remove the SHARP optics cover assembly, then proceed to the next
step below. Refer to “
Removing the SHARP Optics Assembly
”
described previously,
4.
Disconnect the plumbing connections at the rear of the measurement
case assembly.
5.
Disconnect the connectors that pass through the center of the partition
panel.
6.
Remove two screws from the left side rear of the case (viewed from the
front).
7.
Remove one screw from the bottom rear of the case.
8.
Remove one screw from the top front of the partition panel.
9.
While holding the case securely, loosen the captive screw at the rear of
the measurement case assembly, and pull the measurement case
assembly from the rear of the case.
10.
Remove the screw at the top rear of the partition panel that secures the
top of the panel to the measurement case assembly, and lower the
panel, being careful not to put excessive tension on the cables.
11.
Replace the measurement case assembly by following the previous steps
in reverse.
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......