Operation
Service Menu
3-88
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
The Calibrate Barometer Pressure Offset screen allows the user to calibrate
the pressure sensor at offset pressure.
Note
The operator should use an independent barometer to measure the
ambient pressure and enter the value on this screen before calibrating.
▲
●
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration >
Offset
.
CAL BARO PRES OFFSET:
PRESSURE: 760 mmHg
OFFSET: 74
7
mmHg?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Calibrate Barometer Pressure Span screen allows the user to view and
pressure sensor calibration span point.
Note
The operator should use an independent barometer to measure the
ambient pressure and enter the value on this screen before calibrating.
▲
●
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration >
Span
.
CAL BARO PRESSURE SPAN:
PRESSURE: 760 mmHg
SET TO: 74
7
mmHg?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Restore Default Calibration screen allows the user to reset the default
values.
●
In the Main Menu, choose Service > Pres/Vacuum Calibration >
Baro Pres Calibration >
Set Defaults
.
Calibrate Barometer
Pressure Offset
Calibrate Barometer
Pressure Span
Restore Default
Calibration
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......