The Manual Flow Calibration screen allows the user to view and
incrementally set the flow sensor calibration span factor.
●
In the Main Menu, choose Service > Flow Calibration >
Manual
.
FLOW MANUAL CALIBRATION:
FLOW: 16.67
SPAN: 0.750 ?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Mass Calibration submenu is used to view and set the mass calibration
point. The mass calibration screen is visible only when the instrument is in
service mode. For more information on the service mode, see “
Service
Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Mass Calibration
.
MASS CALIBRATION:
>THERMAL COEFFICIENT
VACUUM COEFFICIENT
BARO COEFFICIENT
MASS COEFFICIENT
ALPHA EFF COEFFICIENT
RANGE AVG DIAGS ALARM
The Thermal Mass Coefficient screen allows the user to view and set the
thermal coefficient using manual calibration mode. (Factory set.)
●
In the Main Menu, choose Service > Mass Calibration >
Thermal
Coefficient
.
Manual Flow Calibration
Mass Calibration
Thermal Mass Coefficient