C-Link Protocol Commands
Diagnostics
B-26
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
Send:
io
bd
Receive:
io bd yes
mass
This command reports the presence of mass. The following example
reports that the mass is 0.1 mg/m
3
.
Send:
mass
Receive: mass
0.100E+00
neph bkg
This command reports the current nephelometer background values stored
on the nephelometer interface board. The following example reports the
nephelometer auto-ranging background values.
Send:
neph
bkg
Receive:
neph bkg1 0.00272 0.00272 0.01099 0.04351 0.17430
0.65535
neph mode
This command reports the current operating mode of the nephelometer.
The following example shows that the nephelometer is in sample mode.
Send:
neph
mode
Receive: neph
mode
sample
neph prev bkg
This command reports the previous nephelometer background values
stored on the nephelometer interface board. The following example reports
the previous nephelometer auto-ranging background values.
Send:
neph
prev
bkg
Receive:
neph prev bkg 0.00272 0.00272 0.01094 0.04357 0.17422
0.65535
set reset tape counter
This command resets the tape count to zero.
Send:
set reset tape counter
Receive:
set reset tape counter ok
sample rh
This command reports the sample RH, in percent. The following example
reports that the sample RH is 20%.
Send:
sample
rh
Receive:
sample rh 20.0 %
Summary of Contents for 5030i SHARP
Page 4: ......
Page 22: ...Contents xviii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 26: ...Figures xxii Model 5030i SHARP Instruction Manual Thermo Fisher Scientific ...
Page 168: ......
Page 218: ......
Page 294: ......
Page 306: ......
Page 402: ......
Page 403: ......