Operation
Service Menu
Thermo Fisher Scientific
Model 5030
i
SHARP Instruction Manual
3-91
●
●
●
●
The Flow Calibration submenu is used to view and set the flow calibration
flow point. The flow calibration screen is visible only when the instrument
is in service mode. For more information on the service mode, see “
Service
Mode
” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
▲
●
In the Main Menu, choose Service >
Flow Calibration
.
FLOW CALIBRATION:
>AUTO
MANUAL
RANGE AVG DIAGS ALARM
The Auto Flow Calibration screen allows the user to view and set the
correct flow rate. The span value is automatically calculated.
●
In the Main Menu, choose Service > Flow Calibration >
Auto
.
FLOW AUTO CALIBRATION:
FLOW: 16.67 LPM
SET TO: 17.2
5
LPM?
MOVE CURSOR
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Flow Calibration
Auto Flow Calibration
RESTORE DEFAULT CAL:
RESTORE
RANGE AVG DIAGS ALARM
RESTORE DEFAULT CAL:
RESTORE
ARE YOU SURE YOU WANT TO?
x
PRESS
TO CONFIRM RESTORE
RANGE AVG DIAGS ALARM