Operation
Alarms Menu
3-78
Model 5030
i
SHARP Instruction Manual
Thermo Fisher Scientific
The Flow screen displays the current flow reading and sets the minimum
and maximum alarm limits. Acceptable alarm limits range from -5 mmHg
to 40 mmHg. If the flow reading goes beyond the minimum or maximum
limit, an alarm is activated and the alarm (bell) icon appears in the status
bar on the Run screen and in the Main Menu.
●
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms >
Flow
.
FLOW:
ACTUAL 23.1 mmHg
>MIN -5.0 mmHg
MAX 40.0 mmHg
RANGE AVG DIAGS ALARM
The Minimum Flow alarm limit screen is used to change the minimum
flow alarm limit. The minimum and maximum flow screens function the
same way.
●
In the Main Menu, choose Alarms > Pressure/Vacuum Alarms >
Flow >
Min
or
Max
.
FLOW PRESSURE:
ACTUAL MIN: -5.0 mmHg
SET MIN TO: -4.9 mmHg?
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Flow Alarms submenu allows the user to view and set the flow alarm
parameters.
●
In the Main Menu, choose Alarms >
Flow Alarms
.
Flow
Min and Max Flow Limits
Flow Alarms